Conductivity
Digital sensor for conductivity.
Digital sensor for measuring conductivity, especially in pure media, for use with TriBox controllers and HS100 DIN G2 rail modules. The digital technology ensures secure and interference free signal transmission from the sensor to the controller. |
Measurement technology | Conductivity | |
Measurement principle | Conductivity with two graphite electrodes | |
Parameter | Conductivity | |
Measurement range | 0.00 µS… 20000 µS | |
Measurement accuracy | ±0.5 µS at 20 µS ± 5 µS at 200 µS ± 50 µS at 2000 µS ± 500 µS at 20000 µS | |
Response time | T90 < 60s | |
Temperature compensation | Via NTC | |
Housing material | PVC housing, graphite electrodes | |
Dimensions (L x Ø) | 220 mm x 33 mm | ~ 8.7“ x 1.3“ |
Interface | RS485 (Modbus RTU) | |
Power supply | 12…24 VDC | |
Connection | 8-pin M12 connector, cable length 2 m or 10 m | |
Maintenance interval | 2 years | |
System compatibility | Modbus RTU | |
Warranty | 1 year (EU & US: 2 years) on electronics; wear parts are excluded from the warranty | |
Process pressure | 10 bar | ~ 145 psig |
Calibration method | One-point calibration with standard measuring solution | |
Process temperature | 0…50°C | ~ +32 °F to +122 °F |
![]() |